FX-1000p-B Bufferless Chlorine Analyzer "Mounting and Location"
Foxcroft Equipment and Service Co., Inc. 2101 Creek Road Glen Moore, PA 19343 USA 610-942-2888 FAX 610-942-2769
email sales@foxcroft.com
|
|
Location: Mount the FX-1000p-B Bufferless Amperometric Chlorine Analyzer in a location that is:
Protected from the elements (rain, snow, ice, dust, ect.)
Clean and dray where there is little or no vibration. Protected from corrosive fluids.
Within ambient temperature (32-120 degree F, 0-40 degree C)
Important Mounting the system in direct sunlight may increase the ambient temperature above the maximum limit.
Mounting:
Attach the four mounting ears to the rear of the remote enclosure. Anchor securely to a wall with the sample flow cell drain outlets pointing down. Mount the FX-1000p-B level to ensure that its integral weir operates properly, and at approximate shoulder height for convenience. |
||||||||||||||||||||||