FX-1000p-B Bufferless Chlorine Analyzer "Electrical"
Foxcroft Equipment and Service Co., Inc. 2101 Creek Road Glen Moore, PA 19343 USA 610-942-2888 FAX 610-942-2769
email sales@foxcroft.com
|
Conduit Holes: Run all wiring to the FX-1000p-B Bufferless in 1/2-inch, grounded metal conduit. If using only shielded cable, appropriate strain relief or cable grips are required. Seal unused cable entry holes with appropriate plugs.
Important The electronic enclosure is not a junction box and should not have any other wiring running through it.
This non-metallic enclosure does no automatically provide grounding between the conduit connections Grounding must be provided as part of the installation.
Electrical Connections: Make electrical connections to terminal strips located on the bottom circuit board in the enclosure. To access this board, unfasten the retaining nuts on the four threaded standoffs, and carefully remove the cover plate and top display/controls circuit board. Make sure that the ribbon interconnect cable remains securely attached at both ends. |
|